Home  >  Product  >  Sterilization Indicator Series  > Sterilization Indicator Tape

Sterilization Indicator Tape

 

 

1.  Pressure Steam Sterilization Tape

(1) Usage scope: It mainly applies to indicate whether 

various pressure steam sterilizer for

 fixing the pack and indicating whether the pack(container) 

has been sterilized.

(2) Usage: Intercept proper length tape, paste it to the 

surface of pending sterilized package 

and then sterilize. After sterilization, pick out the

 package and observe the change of tape. 

If the color changes from light yellow to grey/black,

then the pack has been sterilized.

(3) Note: The items to be sterilized should not be placed close to interior wall of the cabinet, 

the tape should not attach to the metal that is easy to form condensation and other hard surface

 such as glass etc, so as to avoid accuracy missing of color agent during the sterilization.

(4) Storage: away from light , corrosive gas and in 15℃~30℃, 50% relative humidity.

(5) Validity: 24months after producing.

 

2.  Autoclave Sterilization Tape for EO Sterilization 

(1) Usage Scope: For chemical monitoring of EO sterilization to check whether the item has been

 sterilized, Fit for 100%EO and EO/HCFC gas mixture sterilizer.

(2) Usage: intercept proper long tape, attach it to the pending items and observe the color changing, 

which indicating whether the items have passed the EO sterilization.

(3) Note: only for chemical monitoring of EO sterilization, not for steam or dry heat sterilization monitoring.

(4) Storage: away from light , corrosive gas and in 15℃~30℃, 50% relative humidity.

(5) Validity: 18 months after producing.

 

3.  Plasma Sterilization Instructions Tape

(1) Usage scope: Indication of hydrogen peroxide plasma sterilization under low temperature

(2) Usage: 1. Intercept proper length tape, paste it to the surface of pending sterilized package 

and then sterilize. After sterilization, pick out the package and observe the change of tape. 

If the color changes to yellow from initial green, then the pack has been sterilized.

(3) Note: The items to be sterilized should not be placed close to interior wall of the cabinet, 

the tape should not attach to the metal that is easy to form condensation and other hard 

surface such as glass etc, so as to avoid accuracy missing of color agent during the sterilization.

(4) Storage: in 15℃~30℃, 50% relative humidity, away from light and corrosive gas.

(5) Validity: 18 months after producing.

 

4.  Formaldehyde Sterilization Instructions Tape

(1) Usage Scope: Sterilization monitoring of formaldehyde sterilizer.

(2) Usage:  Intercept proper length tape, paste it to the surface of pending sterilized package 

and then sterilize. After sterilization, pick out the package and observe the change of tape.

 If the color changes to light green from initial red, then the pack has been sterilized.

(3) Note: The items to be sterilized should not be placed close to interior wall of the cabinet, 

the tape should not attach to the metal that is easy to form condensation and other hard 

surface such as glass etc, so as to avoid accuracy missing of color agent during the sterilization.

(4) Storage: in15℃~30℃, 50% relative humidity, away from light and corrosive gas.

(5) Validity: 12 months after producing.