B-D Test Package

B-D Test Package

Text: [Big] [middle][small] Release date: 2013-11-23  Views: 1887

 

 

Usage scope : Specialized in monitoring whether the cold air

was completely exhausted from forevacuum

or pulsation vacuumpressure steam sterilizer or stuff packs .

Besides , the existence of cold air can be also

 inspected . The color of test chart puncturedby steam with cold

air can be also inspected . The color of

test chart punctured by steam with cold air is hard to fade after

sterilization . Usage : Inspect whether the

 cold air in sterilizer was exhausted effectively after pumping .

The test bag should be used directly under

132℃-134℃ for 3.5min without workers packaging .

Judgment : The color of test chart will turn regular black from initial light yellow after a cycle sterilization

 under 132℃-134℃ for 3.5min .

Storage : room temperature (15℃-30℃)and dry area( humidity:<50%) as well as free from light

Validity : 24months after producing .

B-D test chart is bottom on paper, the size is the same as the cross-section of standard test package, and the front of paper is printed with a light yellow bar pattern,

the color after sterilization will not fade.

B-D test chart is the steam penetration test chart ,that dedicated to testing in pre-vacuum or pulsation vacuum pressure steam sterilizer or the packaging,whether the cold air is

completely discharged,and whether there is cold air in the package.

Used to detect that the air will be effectively eliminated from the sterilizer when prevacuum steam

sterilizer phase in vacuum exhaust stage,the use conditions is that in temperature of 132°C ~ 134°C

keeping the role of 3.5 to 4.0 minutes, test packaging can be used directly, and the staff don’t do standard package.

 Processed through a sterilization cycle,in temperature of 132°C ~ 134°C,

keeping the role of 3.5 min, the light yellow bar patterns of test chart should be turned to the uniform black color.

Storage condition:Keep stored at room temperature 15°C ~ 30°C and dry place (<50% relative humidity) ,avoid to

directly contact with sunlight.

1. The saturated stream pressure should reach to 0.4Mpa 0.6Mpa if air is supplied through

 pipe; and nonuniform color changes may be observed if the pressure is lower than 0.3Mpa;

2. As for the pre-vacuum steam sterilizers or high vacuum autoclave with pulsing systems

 with air supply by pipe, the time of stoppingair supply should longer than 2 days; and airremoval

process should be done before reusing, or else, large amount of aqueous vapor existing in saturated stream may cause difficulty in cool air removal and wet pack, which will result

in incapability of demonstrating test results due to wet BD test paper.

Since the production date 24 months.

Produced and packed by the customer’s request .